When we redo the test with QAWeb Agent rendered test patterns, then the test passes.
We regularly come across failed DICOM compliance tests where the test patterns were measured in PACS.
Make sure to attach the sensor to the workstation and wait for the driver to load correctly before opening QAWeb Agent. You can find the full list in the QAWeb Agent user guide which is linked to this KB article (scroll to the bottom of the page). Several external sensors are supported in QAWeb Agent. For your convenience, we have added such a sheet to this article.
When using an unsupported sensor, then you will need an external calculation sheet. The benefit is that your test results are automatically available and are stored in QAWeb for future reference. When you use a QAWeb supported sensor, QAWeb can do the calculations automatically for you. A tool to calculate/evaluate the measurements. Stable ambient light in the room which is in line with the calibrated ambient light settings. TG18-LN12 test patterns, preferably using QAWeb Agent to display the test patterns (see later). when doing the test on a Coronis Uniti MDMC-12133, you need a sensor which is capable of accurately measuring 1000 Cd/m 2) A calibrated external sensor which has a luminance range which exceeds the range of the display (e.g. What do you need for correctly performing a DICOM compliance test with an external sensor? A visual test on your display failed, even though the DICOM compliance test with the internal sensor passed. Your own QA system requires each sensor to be verified periodically. You question the accuracy of the built-in sensor. You have a regulation in your country or region which requires this (e.g. However, in some cases, there is a need for a test with an external sensor. This feature is very convenient and saves you a lot of time. Why perform a DICOM compliance test with an external sensor?Īll currently selling Barco diagnostic displays have a built-in sensor which is capable of doing an intervention-free DICOM compliance test. How to perform a DICOM compliance test with an external sensor and QAWeb Agent.Distance measurement versus on-screen measurement.Test patterns in QAWeb Agent versus PACS.Why perform a DICOM compliance test with an external sensor?.This article applies to the following products: